Gas Analyzers for the R&D of Carbon Neutral Emissions
A&D offers a variety of in-cell and in-vehicle Gas Analyzers for emissions testing including testers that detect H2, NH3, and our BEX-1000FT multi-component gas analyzer based on Fourier Transform Infrared (FTIR) technology. This tester allows for the continuous raw or dilute measurements of up to 40 gases using one device. If you’re looking to implement emissions measurement or a user-friendly emissions analyzer / gas analyzer, look no further than BEX-1000FT.
A&D also has several Hydrogen Emissions Analyzer Systems. Learn more.
In-Cell Emission Measurement and Testing
BEX-1000FT, a multi-component emissions analyzer based on Fourier Transform Infrared (FTIR) technology, allows for the continuous raw or dilute measurements of up to 40 gases using one device, which allows the system to replace multiple conventional analyzers. Since there is only one detector, there is only one time delay to handle –compared to multiple time delays when using numerous conventional analyzers. BEX-1000FT is a standalone option for a FTIR gas analyzer and key piece of emissions measurement equipment. The standard system operates at vacuum pressure which allows for a fast-response time and lower consumption of the exhaust gas. It can also operate at atmospheric pressure, thus providing a more tailored approach to sample acquisition.
- Provides asynchronous measurement of over 30 regulated and unregulated gases at one detector
- Vacuum sampling system that allows for decreased absorption, a fast response rate (T10-90 -approximately 1.5 sec) and a lower consumption of the exhaust sample
- One device that replaces multiple conventional analyzers, with only one time delay to consider for dynamic analysis
- Optional roll cabinet available to provide easy sharing between test cells
- Designed to provide continuous measurement of undiluted (raw) vehicle exhaust for gasoline, diesel, CNG and alternative fuel vehicles
- Low-cost operation due to drift-free calibrations and minimal maintenance requirements
- A single device that measures all gasses imperative for emissions regulations as well as engine and catalyst calibration and development
A&D offers a variety of hydrogen emissions analyzer systems. Learn more about them here.
- Differentiated measurement of NO, NO2, N2O, CO, CO2, CH4, NH3, hydrocarbons, alcohols and carbonyls
- Sample acquisition rate up to 5Hz
- An advanced cross-interference compensation routine, significantly minimizing the influence of water and other interfering gasses
- High Pressure Control system, allowing for sampling in high-pressure, sample-inlet conditions
- Meets or exceeds all requirements specified for FTIR Measurements by US EPA, ASTM, ISO and Euro VI HDD
- Vacuum sampling method allows for a faster response rate (T10-90<1.5 sec), and a lower consumption of the exhaust sample
- Easy integration into automation systems via AK interface
- Transportable analyzer that allows for test cell sharing
- Gas analysis methods for different fuel types, including gasoline, diesel, CNG, and other alternative fuels
- Low-cost ownership due to drift-free calibration and minimal maintenance requirements
- One physical analyzer with one delay time to simplify the measurement process
- Excellent correlation with conventional analyzer systems
BEX-1000FT SPECIFICATIONS |
|
---|---|
Measuring Method | FTIR |
Measuring Component | 30 Components standard, up to 40 optional |
Application | Gasoline/Diesel Direct |
Dimensions | Approximately 550(W) x 800(D) x 1250(H) |
Weight | Approximately 80 kg |
Operating Conditions | 5-40°C, Relative Humidity 80%(non-condensing) |
Sampling Frequency | 1 Hz and 5 Hz (selectable) |
Absorbance Spectrum | 500-5000 cm-1 |
Spectral Resolution | 0.5 cm-1 |
Optical Path Length | 5.1 m |
Gas Cell Volume | 200 ml |
Gas Cell Temperature | 191°C |
Detector Cooling Method | MCT Liquid Nitrogen, 50 ml/h |
SYSTEM PERFORMANCE |
|
---|---|
Response Time (T10-T90) | 1.5 -2.0 sec |
Zero Drift | 1.0% F.S./4 hrs=(±5°C) |
Span Drift | 1.0% F.S./4 hrs=(±5°C) |
Repeatability | 2.0% F.S. |
For a complete list of specifications, including those for fast response time, cross-interference compensation, and configurable high and low ranges, download the following document:
Related Products
iTest Application
The iTest Application offers a wide range of functionality for Test Automation. This family of applications include: iTest Battery, iTest Pro, iTest Standard, iTest Micro and iTest Racing. iTest will configure to any application!
In-Vehicle Emissions Testing
The BOB-1000FT is the only FTIR system optimized for use on-board, and includes integration of data from the ECU, weather data, GPS data and time-aligned video.
Lab Management Tools
LabWorX Suite of products will transform a collection of test systems into one managed lab.
iTest Application
The iTest Application offers a wide range of functionality for Test Automation. This family of applications include: iTest Battery, iTest Pro, iTest Standard, iTest Micro and iTest Racing. iTest will configure to any application!
In-Vehicle Emissions Testing
The BOB-1000FT is the only FTIR system optimized for use on-board, and includes integration of data from the ECU, weather data, GPS data and time-aligned video.
Lab Management Tools
LabWorX Suite of products will transform a collection of test systems into one managed lab.
Get In Touch
United States | Ann Arbor, MI
SALES: (734) 822-9502
SUPPORT: (734) 822-9556
Location
4622 Runway Blvd
Ann Arbor, MI 48108
contact@aanddtech.com
A&D Europe | United Kingdom
Location
24 Blacklands Way
Abingdon, Oxfordshire OX14 1DY United Kingdom
Phone:
+44 (1235) 55 04 20 Ext 131
Email:
ptservice@aanddeurope.com